​​​​​​​​​C‑Flow PFA Concentric Nebulizer

The C-Flow is a molded, fixed-capillary st​​​yle PFA (perfluoroalkoxy) concentric nebulizer for the introduction of low-volume (<1 mL)​ s​​​​am​​​​ples to the Teledyne CETAC Aridus3 Desolvating Nebulizer System. The Aridus3 is used for signal enhancement / interference reduction with ICP-MS.

Inert Construction: The inert, fluoropolymer construction of the C-Flow PFA Nebulizer allows introduction of all acids (including hydrofluoric acid), alkalis, and organic solvents.

Standard Size: The C-Flow features a standard 6 mm diameter tip for easy installation to the Aridus3 heated PFA spray chamber. The nebulizer gas supply is from the host ICP-MS instrument, with 4 mm diameter tubing attaching directly to the C-Flow nebulizer side gas inlet port.

Optimized for the Aridus 3/ASX-112FR C-Flow PFA nebulizers for the Aridus3​ are factory adjusted for best performance in the heated PFA spray chamber and feature a built-in autosampler probe for connection to the ASX-112FR MicroAutosampler.

Low-Flow Rates Available: The C-Flow PFA Nebuliz​er for the Aridus3 is available in three self-aspiration flow rates of 50 µl/min, 100 ​µL/min, and 200 µL/min.​​ The 100 µL/min​ nebulizer is included in the Aridus3 One Year Consumable Kit​.

 Specifications

Speci​fication Title Specification Details
Sample Uptake Rates
50µL/min, | 100µL/min | 200µL/min
Nebulizer Gas Flow Range 0.8 to 1.2 L/min
Nebulizer End (Outer Diameter) 6 mm 
Nebulizer Overall Length 820 mm (50, 200 µ​L/min versions)
1150 mm (100 µ​L/min version)
Nebulizer Uptake Line Length 745 mm (50, 200 L/min versions) 
1075 mm (100 µL/min version)
​​Warranty ​​12 month limited​​​​​​

 Application Notes

  
  
TN-Aridus3-001.pdfMulticollector ICP-MS with a New Desolvating Nebulizer System for U-Th Series Measurements: an Optimization Procedure
PST_Dating-Study-of-Cave_Aridus3.pdfIsotope Ratio Measurements for Sample Dating Studies Using Multicollector ICP-MS and Signal Enhancement via a Desolvating Nebulizer Accessory with Nitrogen Gas Addition
Poster_2018WPC_Aridus3.pdfPoster: System Setup and Parameter Optimization for a Desolvating Nebulizer Unit with Multicollector Inductively Coupled Plasma Mass Spectrometry for Stable and Radiogenic Isotope Ratio Measurements
AN-Aridus3-003.pdfAntimony Isotope Ratio Measurements: A Study with an Atlantic Ocean Mn Nodule
AN-Aridus3-002.pdfSystem Setup and Parameter Optimization for a Desolvating Nebulizer Unit with Multicollector Inductively Coupled Plasma Mass Spectrometry for Stable and Radiogenic Isotope Ratio Measurements
AN-Aridus3-001.pdfUsing Multicollector ICP-MS with a Desolvating Nebulizer Accessory for Stable and Radiogenic Isotope Ratio Measurements of Speleothem and Marine Coral Samples